Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K083514

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-05-22

    (177 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Terumo Syringe with/without needle is a sterile hypodermic syringe for single use, intended for the aspiration of fluids and blood, or for the injection of fluids immediately after filling.

    Device Description

    The Terumo Syringe with/without needle is a hypodermic standard piston syringe, available in volumes from 1 ml, 2 ml, 10 ml and 50 ml, with a luer taper tip for single use, made of plastic material and a synthetic rubber gasket. The needle is made from stainless steel.

    AI/ML Overview

    The provided text describes a 510(k) submission for a medical device, the Terumo® Syringe with/without Needle. It focuses on demonstrating substantial equivalence to predicate devices and adherence to relevant standards, rather than presenting a study of the device's diagnostic performance or algorithm effectiveness. As such, many of the requested categories for acceptance criteria and study details are not relevant or cannot be extracted from this type of regulatory document.

    However, I can extract information related to the device's performance testing against a recognized standard, which serves as its primary acceptance criteria in this context.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard)Reported Device Performance
    Mechanical Performance: EN ISO 7886-1 (1993) (Sterile hypodermic syringes for single use - Part 1: Syringes for manual use)"The Terumo syringe with/without needle was tested in accordance with EN ISO 7886-1 (1993)." Implicitly, the device meets the performance requirements of this standard for its mechanical and functional aspects. The submission states, "None of the obtained data raises any new issues of safety and effectiveness."
    Needle Properties: (Reference to cleared Neolus Needles, K001572)"For the needle properties, reference is made to the cleared Neolus Needles compared to in the submitted file K001572." This indicates that the needles used with the syringe meet the established properties of the predicate Neolus Needles.
    Hemocompatibility: EN ISO 10993-4 (2002) and amendment (2006) (Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood)"In accordance with the new requirements stipulated in EN ISO 10993-4 (2002) and its amendment (2006) the hypodermic needles were also subjected to additional hemocompatibility testing. A declaration for compliance with the requirements of these tests can be found on page 177." This indicates testing was performed and compliance was declared, implying the device meets these criteria.
    Sterility Assurance Level (SAL): EN 550 ("Sterilization of Medical Devices: Validation and routine control of ethylene oxide sterilization") ISO 11135 ("Medical Devices: Validation and routine control of ethylene oxide sterilization") EN 556-1 ("Sterilization of Medical Devices - Requirements for medical devices to be designated "STERILE" - Part - 1: Requirements for terminally sterilized medical devices")"The sterility of the Terumo Syringe with/without needle is assured by using a validated sterilization method qualified in accordance with EN 550... and ISO 11135... to a sterility assurance level (SAL) of 10^-6 as required by EN 556-1..." This confirms the device achieves the required SAL.
    Ethylene Oxide Residual Levels: ISO 10993-7 ("Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals")"Ethylene oxide residual levels and Ethylene Chlorohydrin residual levels resulting from EtO sterilization are in compliance with ISO 10993-7..." This confirms the device complies with the residual limits.
    Biocompatibility: EN ISO 10993-1 ("Biological Evaluation of Medical Devices. Part-1: Evaluation and testing.")"The device is tested for biocompatibility requirements in accordance with the tests stipulated in EN ISO 10993-1... The results of the testing demonstrate that the device is biocompatible." This confirms the device meets biocompatibility requirements.
    Expiration Dating: (No specific standard mentioned, but a characteristic of medical devices)"The expiration dating for the Terumo Syringe with/without needle has been established at 5 years." This indicates a tested and validated shelf-life.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated. The document mentions "testing" and "data" but does not quantify the sample sizes used for each specific test (e.g., how many syringes were tested for dimensional accuracy, or how many batches for sterility).
    • Data Provenance: The device is manufactured by Terumo Europe N.V., in Belgium. The submission states the device "was tested in accordance with EN ISO 7886-1 (1993)", suggesting the testing was performed, or overseen, by the manufacturer or a certified testing facility in relation to its European operations. The description is of the device's inherent physical and material properties, rather than clinical data from human subjects. It's a laboratory/engineering performance study, not a clinical study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This document describes laboratory and engineering performance testing against established international standards. There is no "ground truth" established by human experts in the sense of clinical diagnoses or interpretations. The "truth" is defined by the objective pass/fail criteria of the scientific/engineering standards.

    4. Adjudication method for the test set:

    • Not applicable. This relates to clinical studies with human review, which is not what this document describes. Performance is assessed against quantitative metrics and methods specified within the referenced ISO/EN standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a manual syringe, not an AI-assisted diagnostic or therapeutic tool. Therefore, MRMC studies and AI performance metrics are irrelevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This device is a manual syringe, not an algorithm.

    7. The type of ground truth used:

    • For the performance testing, the "ground truth" is defined by the objective, quantitative criteria and test methods specified in the referenced international standards (e.g., EN ISO 7886-1, EN ISO 10993 series, EN 550, ISO 11135, EN 556-1). This includes measurements of physical properties, chemical residual levels, and biological responses.

    8. The sample size for the training set:

    • Not applicable. There is no "training set" as this is not an AI/machine learning device.

    9. How the ground truth for the training set was established:

    • Not applicable. There is no "training set" for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K052034

    Validate with FDA (Live)

    Date Cleared
    2005-08-16

    (19 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Terumo syringe with/without needle is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.

    Device Description

    The Iml Terumo Syringe with/without Needle consists of 3 main parts or 4 parts, i.e., a barrel, plunger, gasket and a hypodermic needle (for with needle type). Barrel is made from polypropylene and is designed with clear graduations and figures for easy use. The gasket is made from a highly inert thermoplastic elastomer material. The needle is made from stainless steel. The 1 ml Terumo Syringe with/without Needle is individually packed in peel blister that ensures the sterility of the device until the package is opened. The product is sterilized by electron beam sterilization. The product is for single use only and is not intended for reuse.

    AI/ML Overview

    Here's an analysis of the provided 510k summary regarding the acceptance criteria and the study that proves the device meets those criteria:

    This document is a 510(k) premarket notification for a medical device that demonstrates substantial equivalence to a predicate device, rather than proving performance against novel acceptance criteria through a clinical study. Therefore, many of the requested fields related to AI/MRMC studies or detailed ground truth establishment for novel algorithms are not applicable.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (from Performance Section)Reported Device Performance (from Performance Section)
    Leakage (Aspiration and Injection)Subjected to testing
    Plunger MobilitySubjected to testing
    Plunger Gasket FitSubjected to testing
    Plunger Stopper StrengthSubjected to testing
    DeadspaceSubjected to testing
    Nominal Graduation CapacitySubjected to testing
    Packaging Seal StrengthSubjected to testing
    Biocompatibility (blood contacting materials)Tested in accordance with ISO 10993-1; results demonstrate materials are biocompatible.
    Sterility (SAL)Validated in accordance with ISO 11137-1995 to a SAL of 10⁻⁶.
    Manufacturing ControlsIncluded visual, functional, dimensional, and sterility tests.

    Study Proving Acceptance Criteria:

    The document states that the "1ml Terumo Syringe with/without needle submitted in this premarket notification was subjected to the following tests," listing the performance criteria mentioned above. It also explicitly states, "The 1ml Terumo Syringe with/without needle is substantially equivalent in intended use, materials, design, technology and principles of operation and performance to cleared Terumo Syringes with/without needle cleared under K023271."

    This indicates that the "study" proving acceptance criteria is bench testing for the listed performance characteristics and material biocompatibility/sterility validation as per international standards, followed by a demonstration of substantial equivalence to an already cleared predicate device (K023271). The specific numerical results or detailed methodology for each test are not provided in this summary but would have been part of the full submission.

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes for the bench tests. Data provenance is not applicable in the context of bench testing for a physical medical device; the tests would have been performed in a laboratory setting by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable since this is a physical medical device undergoing bench testing and not an AI/diagnostic device requiring expert ground truth establishment for a test set.

    4. Adjudication method for the test set

    This is not applicable as there is no 'ground truth' in the sense of expert review for diagnostic output. Bench testing results would be objective measurements against defined specifications.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is a manual syringe and does not involve AI or human "readers" in its operation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. This device is a physical, manually operated syringe.

    7. The type of ground truth used

    The "ground truth" for this device's performance is based on predefined engineering specifications and international standards, such as ISO 10993-1 for biocompatibility and ISO 11137-1995 for sterilization. For example, for "deadspace," the ground truth is a specific maximum allowable volume. For "leakage," the ground truth is a pass/fail against a specific pressure or volume.

    8. The sample size for the training set

    This is not applicable as there is no "training set" for a physical medical device like a syringe.

    9. How the ground truth for the training set was established

    This is not applicable as there is no "training set" for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023271

    Validate with FDA (Live)

    Date Cleared
    2002-12-18

    (78 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Terumo Syringe with/without needle is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.

    Device Description

    The Terumo Syringe with/without needle consists mainly of 3 parts or 4 parts: a barrel, a plunger, a gasket and a needle (for the needle type). Barrel is made from polypropylene and is designed with clear graduations and figures for easy use. The gasket is made from a highly inert thermoplastic elastomer material facilitates smooth movement of the plunger and aids the user in controlling dosage. The needle is made from stainless steel and the needlepoint is carefully designed to minimize resistance at the skin surface and aids in smooth, easy insertion. The Terumo Syringe with/without Needle is individually packed in peel blister that ensures the sterility of the device until the package is opened.

    AI/ML Overview

    This document describes the 510(k) summary for the Terumo Syringe with/without Needle. It demonstrates substantial equivalence to a predicate device, K980181, rather than proving performance against specific acceptance criteria in a quantitative study. Therefore, some of the requested information, such as sample sizes, expert qualifications, and specific effect sizes, are not applicable or not provided in the given text.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not present a table of specific quantitative acceptance criteria with corresponding performance metrics. Instead, it lists various tests performed to demonstrate safety and efficacy and asserts that the device is "safe and effective during handling and for the duration of its expiration." The implicit acceptance criterion for each test is that the new device performs acceptably and equivalently to the predicate device.

    Performance TestReported Device Performance
    CleanlinessNot explicitly quantified, but implied to be acceptable.
    Leakage TestNot explicitly quantified, but implied to be acceptable.
    Sliding ResistanceNot explicitly quantified, but implied to be acceptable.
    Plunger and Gasket Fit ForceNot explicitly quantified, but implied to be acceptable.
    Nozzle-needle Fit ForceNot explicitly quantified, but implied to be acceptable.
    Initial Movement ForceNot explicitly quantified, but implied to be acceptable.
    Stopper ForceNot explicitly quantified, but implied to be acceptable.
    Dead SpaceNot explicitly quantified, but implied to be acceptable.
    Equal to Nominal Capacity Graduation ToleranceNot explicitly quantified, but implied to be acceptable.
    Between Main Capacity Graduation ToleranceNot explicitly quantified, but implied to be acceptable.
    Primary Container Seal StrengthNot explicitly quantified, but implied to be acceptable.
    Silicone Amount DeterminationNot explicitly quantified, but implied to be acceptable.
    Elution Test MethodNot explicitly quantified, but implied to be acceptable.
    Analysis of Heavy MetalsNot explicitly quantified, but implied to be acceptable.
    Determination of pHNot explicitly quantified, but implied to be acceptable.
    Sterility Assurance Level (SAL)10⁻⁶ (validated in accordance with ISO 11137)
    Performance (aged and non-aged devices)Aging process does not adversely affect device performance.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for the various performance tests. The data provenance is Terumo (Philippines) Corporation, and the tests were likely conducted prospectively as part of the device development and validation process.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. This is a medical device clearance based on substantial equivalence, not an AI or diagnostic tool requiring expert ground truth for a test set. The "ground truth" here is compliance with established performance standards and equivalence to a predicate device.

    4. Adjudication Method for the Test Set

    Not applicable. This is a medical device clearance based on substantial equivalence, not an assessment requiring adjudication of diagnostic outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is a manual syringe, and its performance evaluation does not involve human readers interpreting diagnostic cases.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. The device is a manual syringe, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is based on:

    • Compliance with recognized standards: For tests like cleanliness, leakage, and sterility (ISO 11137).
    • Engineering specifications and design requirements: For physical and mechanical properties.
    • Demonstrated performance of the legally marketed predicate device: The new device must perform comparably to the predicate.

    8. The Sample Size for the Training Set

    Not applicable. This is a medical device, not a machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1